Biology Atlas

Map / Assays

assay · intermediate · partly written

Isothermal titration calorimetry (ITC)

Isothermal titration calorimetry measures the heat released or absorbed as one purified binding partner is titrated into another, giving affinity and stoichiometry without labels or surface attachment.

Also called: ITC, calorimetric titration

The short version

ITC watches the heat of binding as one partner is dripped into the other. It gives Kd and stoichiometry from the same curve, with nothing labelled and nothing stuck to a surface. It needs more purified protein than most binding assays, works best in a window where c = [M]/Kd is between 1 and 500, and gives no rates. 1

Our inference

How it works more

Each injection releases or absorbs heat in proportion to how much new complex forms. As sites fill, heats shrink; the plot of heat per injection against molar ratio is the isotherm. The midpoint gives stoichiometry, the slope gives Kd and the height gives enthalpy. The quantity c = [M]/Kd decides what can be read: above 10 the curve is clearly sigmoidal; below 10 it blurs and stoichiometry must be fixed; the quoted workable window is 1 to 500 (vendor white paper; the origin of c was not verified). 1

Our inference

Where it breaks more

Too tight (c above about 500) gives a step with no readable Kd; too weak needs impossible protein (10 mM for a 1 mM Kd at c = 10, when many proteins are insoluble above 1 mM). Displacement titrations and fixed stoichiometry are the standard workarounds. The Kd is for purified protein in buffer, so it says nothing about a cell. The source here is a vendor white paper from an instrument seller. 1

Reported, not yet replicated

Price deep

Two university cores publish prices: Scripps $600 for optimisation plus one sample, then $300 per extra sample (undated page), and USC $40 to $80 per hour for PEAQ-ITC time (undated page). Protein production is not included in either rate. No CRO price was found. 23

Reported, not yet replicated

Question answered

How tightly do two purified molecules bind (Kd), in what ratio (stoichiometry), and with what heat (enthalpy), measured in free solution with no label and nothing stuck to a surface? 1

Our inference

Readout physical

Heat. Each injection of one partner into the other releases or absorbs heat in proportion to how much new complex forms. As binding sites fill, the heat per injection shrinks, so the isotherm is a heat-per-injection curve: its midpoint gives stoichiometry, its slope gives Kd and its height gives enthalpy. This is a summary of the standard method, not a quotation. 1

Our inference

Not measured

Kinetics and context. ITC measures equilibrium heat, so it gives no kon or koff in this treatment, and a Kd from purified protein in buffer says nothing about competitors, crowding or modifications in a cell. For very tight binders (c above about 500) the isotherm becomes a step with no readable Kd. 1

Our inference

Setting

Where in cascade

Sample prep

Purified protein partners in matched buffer, enough of them to reach a usable c value. A 1 mM Kd at c = 10 would need 10 mM protein, and the vendor note says many proteins are not soluble above 1 mM, so weak binders are often out of reach. 1

Reported, not yet replicated

Conditions

The shape is set by c = [M]/Kd (roughly, for a 1:1 complex). Above c = 10 the curve is clearly sigmoidal and stoichiometry is measurable; below 10 the inflection blurs and stoichiometry must be fixed. The workable window quoted is 1 < c < 500. Displacement (competition) titrations are used for very tight binders and fixed stoichiometry for weak ones. 1

Reported, not yet replicated

Throughput

One titration at a time; no throughput figure was found in an opened source. 23

Our inference

Cost

Per sample

300 to 600 USD per titration at a university core (first sample with optimisation $600; each further sample $300) (Scripps Biophysics and Biochemistry Core rate card; excludes protein and your time. Academic core rate, page says prices are subject to change and carries no effective date. Not a commercial or contract-lab price.; as of 2026 (page undated))2

Engineerable variants

Displacement (competition) titration

For binders too tight to read directly (c above about 500), a weaker competitor is titrated against the tight one so the apparent Kd falls into the readable window. 1

Reported, not yet replicated

Fixed-stoichiometry fitting

For weak binders at low c, the inflection point is blurred, so the stoichiometry is fixed in the fit rather than read from the curve. 1

Reported, not yet replicated

Data availability

Scripps Biophysics and Biochemistry Core (university rate card) cro

ITC: $600 for optimisation plus one sample, then $300 per extra sample. 2

Caveats. Academic core rate with no effective date; excludes protein; not a commercial price.

Price. 300 to 600 USD per titration (Scripps core rate card, fetched 2026-11-17, undated; as of 2026 (page undated))2

Reported, not yet replicated

USC Nanobiophysics Core (PEAQ-ITC instrument time) cro

PEAQ-ITC instrument time at $40 to $80 per hour, as the page lists it. 3

Caveats. Hourly instrument time, not a per-sample price; page undated; which tier pays which rate was not recorded.

Price. 40 to 80 USD per hour (USC core rate card, fetched 2026-11-17, undated; as of page undated)3

Reported, not yet replicated

No public ITC dataset route found public

The Kastritis benchmark gives Kd values for 144 structurally characterised complexes, spanning about 1e-5 to 1e-14 M, but the abstract was read and its measurement methods were not examined. 4

Caveats. Abstract only; it is a curated benchmark, not an ITC-specific database.

Reported, not yet replicated

Teach this

In one sentence

ITC measures the heat of binding as you titrate one partner into another, giving Kd, stoichiometry and enthalpy without labels. 1

Our inference

The five-minute version

Put one purified protein in the cell, inject the other in small steps and record the heat each time. Early injections give big heats because many sites are free; later ones shrink as sites fill. Fit the curve for Kd, binding ratio and enthalpy. It works when c = [M]/Kd is between about 1 and 500, so very weak binders need huge protein concentrations and very tight binders need displacement titrations. It gives no on or off rates, so pair it with SPR if timing matters. 1

Our inference

What a listener will ask

Why does ITC need so much protein?

To get a readable curve the concentration must be near or above Kd; a 1 mM Kd at c = 10 needs 10 mM protein. 12345

Reported, not yet replicated

Can ITC tell me how long a complex lasts?

Not in this treatment: it measures equilibrium heat, so kon and koff come from methods such as SPR.

Does the Kd hold in a cell?

No source here quantifies that; it is a purified-protein number and competitors, crowding and modifications are absent.

The common wrong idea

A tighter Kd is always easier to measure.

Above c about 500 the ITC curve becomes a step with no readable Kd, so the tightest binders need displacement titrations. 12345

Reported, not yet replicated

The one citation to hand over: 1

Still open here

Not yet written: Time per run, Use cases, Before after, Capital, ML role.

Linked from

Sources

  1. Studying low-affinity fragments or ligands by ITC (white paper). Malvern Panalytical. vendor-page, not peer reviewed; opened 2026-11-17 all records citing this
  2. Scripps Biophysics and Biochemistry Core: services and prices. Scripps Research (2026). organisation-page, not peer reviewed; opened 2026-11-17 all records citing this
  3. USC Nanobiophysics Core: rates. University of Southern California Dornsife. organisation-page, not peer reviewed; opened 2026-11-17 all records citing this
  4. Kastritis, P. L., et al.. A structure-based benchmark for protein-protein binding affinity. Protein Science (2011). paper; opened 2026-11-17 all records citing this
  5. Oxford Dunn School of Pathology. SPR guidelines (Sir William Dunn School of Pathology, University of Oxford). University of Oxford (2001). web, not peer reviewed; opened 2026-11-17 all records citing this
Where it sits on the map
History of this record (2)

This record has been changed 2 times, on the dates below. The note kept against each change is written for the people who maintain the atlas, in their own shorthand, so it is not shown here; it is in the record itself.

Throughput class
low
Emerging
no
Status
partly written
Last updated
2026-11-17
Record
content/assays/isothermal-titration-calorimetry.json